Validation of a flow cytometry-based assay for monitoring complement-dependent cytotoxicity of an anti-CD20 biosimilar candidate antibody

Author
Rancés Blanco Santana, Mercedes Cedeño Arias, Zulia García Pérez, Dayana Columbié Gilbert, Alejandro Portillo Vaquer, Enrique Rengifo Calzado
Date

 

Abstract

Appropriate validation of any bioassay to be used in the characterization of biological products is critical. In this study, we report a validation study of a flow cytometry-based assay to measure the complement-dependent cytotoxicity (CDC) of a biosimilar candidate monoclonal antibody (Mab) directed to CD20 antigen, as indicative of its biological activity. The assay was validated by examining: assay robustness, specificity, repeatability and intermediate precision. It demonstrated to be robust for all factors evaluated. It also showed a high level of specificity and was found to be free of interference through the validation process. The degree of precision (Cvs < 7%) obtained in this study was satisfactory. The presented work demonstrated that a flow cytometry-based cytotoxicity assay is a suitable method in the lot to lot quality control monitoring of the culture supernatant and an active pharmaceutical ingredient of 1B8 Mab.

Key words: flow cytometry, complement-dependent cytotoxicity, validation study, therapeutic anti-CD20 Mab, Biosimilar